BM – Moisture Balance

QUICK OVERVIEW

Product Code:543
Brand:Balance

BM Series


Share this product on social networks

BM Series Moisture Analyser With halogen heating is mainly designed on the basis of time saving factor for universal use in chemical, pharmaceutical, oil, petroleum industries , material testing, environmental Laboratories, and sewage treatment plants. TheAcculab’s MB Series Moisture Analyzer is compact and big in performance. It performs all the major Functions of routine moisture analysis.

Features:

  • Large, easy to read graphical LCD display with AEP.
  • Memory for 99 users’ selectable procedure for analysis.
  • 4 different ways of drying.
  • User selected visualization display for process of drying.
  • Temperature range up to 106 C
  • Procedures of GLP.
  • Automatic finish of procedure for drying.
  • Declaration of sample.
  • Standard printouts.
  • Easy programmed printouts.
  • Improved repeatability for better results

Technical Specification

MODEL

BM 53

BM120

BM200

BM 54

                                                                             Measurement Value

Readability

0.01%

0.01%

0.01%

0.001%

Readability (sd) for initial sample wt.= 2g

0.1%

0.1%

0.1%

0.05%

Readability (sd) for initial sample wt.= 10g

0.03%

0.03%

0.03%

0.02%

                         Evaluation

Units of Measurement

% M, %D, %R (ratio), Gm/Lt, %M Atro, %D Atro

Statistics

Method wise for each individual 40 methods or
Batch wise for each individual 100 batches for last 9999 results

Data Storage

Last 1000 Moisture Determination results

                          Weighing Balance

Max. Sample weight

50 g

120g

200g

50 g

Minimum Sample Weight

0.2 g

0.2 g

0.2 g

0.2 g

Readability

0.001 g

0.001 g

0.001 g

0.1mg

                           Drying Unit

Technology

Infra Red Halogen

Temperature range

30º C to 200º C

Temperature increment

1º C

Heating Profile

Standard, Gentle, Rapid, Steps, High Temperature

Swtich off Criteria

Auto, Manual, User Def Weight / Time, User Def % unit / time, Intelligent

                         User Support

Auto Method Development

Intelligent

Method Memory

40

Batch Memory

100

Method Designation

Alpha Numeric

Serial Number

YES

Compile

Yes (Method Linking)

Standby Temperature

30 C to 100 C

Display

Graphical LCD display with LED backlit

Data Interface

RS232C Bidirectional, USB, Ethernet, PS2

Print

GLP / GMP Compliant, User Configurable

                        Quality Management

Test

Weighing and Heater

Weighing in Aid

YES

Password

Method, Setting, Print Format, Data Delete, Calibration, Drying

                  Miscellaneous

Power Supply

230 v/50Hz or 115v/60Hz

Power Consumption (Watt)

Max 415

Housing Dimension mm

211 (w) x 342 (d) x 187 (h)

Weight Approx

5kg

Packing Dimension mm

325 (w) x 390 (d) x 285 (h)

LITERATURE