MPM 5000A

QUICK OVERVIEW

Product Code: 170062
Brand: Multi Parameter

MPM5000A


Share this product on social networks

MPM 5000A multi-parameter online analyzer: suitable for long-term on-line online monitoring, measuring up to 8 parameters simultaneously, including: Temperature, Depth, pH, ORP, Conductivity & Salinity & TDS, Turbidity, DO, Chlorophyll,

Blue-green algae, Ammonia nitrogen, Nitrate, Chloride, Fluoride.

Optional accessories such as bluetooth module, battery pack, and handheld communicator are also available for easy portable measurement.

The Communicator: In order to make the MPM5000A integrated multi-parameter analyzer complete data reading,

parameter setting and calibration of the analyzer without PC, and it can be applied to handheld and portable applications. On the occasion, the company has developed a HDC-100A handheld device for use with it. Built-in GPS receiver; with bluetooth function; with USB interface and data storage function; designed with ergonomic curve, rubber gasket, suitable for hand-held operation, easy to grasp in wet environment.

  1. The host can flexibly set the sampling interval, optimize the working/sleep time, and reduce the overall power consumption. It can be applied to battery power, solar power, and mains power supply
  2. SensorRS485 A/B terminal is connected to power supply
  3. The host can select blue-tooth communication function to realize data transmission via blue-
  4. Optional handheld communicator, and blue-tooth wireless communication mode with the handheld communicator, data viewing, calibration is convenient, and can realize handheld
  5. The host has a data storage function, and more than 100,000
  6. Self-cleaning system allows the product to obtain accurate data for a long
  7. he electrode can be replaced directly in the field, and the maintenance is simple and
  8. Sensor power supply positive and negative reverse connection
  9. The host has LED lights to display the current running status of the host and the working status of blue-tooth.
  10. StandardModbusRS-485 digital signal output, can be connected with third-party data loggers or remote devices for integration and networking; with built-in battery, optional battery to

Software Fun Functions:

  • Operation software of windows interface has the function of settings, online monitoring, calibration and historical data download, convenient and efficient parameters Convenient and efficient calibration functions.
  • Real-time data and curve display can help users intuitively obtain the data of measured water
  • Intuitively and accurately understanding and tracking the parameters’ changes of the measured water bodies in a certain period through the Historical data download and curve display.

 

  • Application:
  • Multi-Parameter water quality online monitoring of rivers, lakes and reservoirs.
  • Water quality online monitoring of drinking water source, ground water and sea water.

 

TECHNICAL SPECIFICATION

Mainframe Physical Indicators

Power Supply

 

9-36VDC

Measuring Temperature

0-40℃(non-freezing)

Power Dissipation

3W

Storage Temperature

0-50℃(non-freezing)

Communication protocol

MODBUS RS485

Size

Diameter 90*Length 635mm

Maximum withstand pressure depth

30M(With ion selective electrode: 10 m)

Protection Class

IP68

weight

5Kg

Battery Capacity

8 sections, 3.6V lithium subbattery

Communicator Physical Indicators

Display

 

3.5-inch color display screen with adjustable backlight

Data storage

More than 100,000 data

Material

ABS+PC

Power supply                                              Built-in battery power, battery specifications: 4 3.7V rechargeable lithium battery

Protection level

IP67

Operating & Storage temperature

0-50℃(Non-freezing),-15-60℃

Size &Weight

203*100*43mm , 0.5KG

 

Standard Electrode Parameters :
pH Sensor :
Principle

Glass electrode method

Range

0-14 pH (With ion electrode:5- 10 pH)

Resolution

0.01 pH

Accuracy

±0.1 pH

Conductivity sensor:
Principle

A pair of platinum gauze electrode

Range

1us/cm-2000 us/cm (K=1)&100us/cm-100ms/cm (K=10.0)

Resolution

0.1us/om-0.01ms/cm (Depending on the range)

Accuracy

±3%

Do Sensor:
Principle

Fluorescence

Range

0-20 mg/L: 0-20 ppm:  0-200%

Resolution

0.1%/0.01mg/l

Accuracy

±3% or ±0.3 mg/L, whichever is greater

Chlorophyll Sensor:
Principle

Fluorescence

Range

0-500 ug/L

Resolution

0.01 ug/L-0.1 ug/L, Depending on range

Accuracy

±5% of the signal level corresponding & value of 1 ppb Rhodamine B Dye

Ammonia Sensor:
Principle

long selection

Range

0.1-3000mg/L-N

Resolution

0.01mg/L-N

Accuracy

±5% or ±0.2mg/L

Chloride Sensor:
Principle

long selection

Range

3-3500 mg/L

Resolution

0.01mg/L

Accuracy

±10% or±0.2mg/L

Depth (Pressure):
Principle

Pressure-sensitive Method

Range

0-61m(Maximum depth:30M) withstand pressure

Resolution

2cm

Accuracy

±0.3%

Temperature:
Principle

Thermistor method

Range

O°C-60°C (Mainframe’s work temp. 0~40°C)

Resolution

0.1°C

Accuracy

±0.51C

Standard Electrode Parameters :
ORP Sensor:
Principle

Glass electrode method

Range

-1000mV-+1000mV

Resolution

1mV

Accuracy

±5mV

Turbidity Sensor:
Principle

Light scattering method

Range

0-1000 NTU

Resolution

0.1NTU

Accuracy

± 5% or 0.3NTU, whichever is greater

Blue-green algae Sensor:
Principle

Fluorescence

Range

0-300,000 cells/mL

Resolution

20 cells/mL

Accuracy

±10% of the signal level corresponding value of 1 ppb Rhodamine B Dye

Nitrate Sensor:
Principle

long selection

Range

0.5-3000mg/L-N

Resolution

0.01mg/L-N

Accuracy

±10% or ±0.2mg/L

Fluoride Sensor:
Principle

long selection

Range

0.5-10000 mg/L

Resolution

0.01mg/L

Accuracy

±10% or ±0.2mg/L

Potassium Ion Sensor:
Principle

long selection

Range

0.5-5000 mg/L

Resolution

0.01mg/L

Accuracy

±10% or ±0.2mg/L